Air Velocity Transducer Models 8455, 84659 8475
<br />8455 *000 8465 8475
<br />Accuracy7: ±2.0% of readingl ±2.0% of readingl ±3.0% of readingl
<br />±0.5% of full scale of selected range ±0.5% of full scale of selected range ±1.0% of full scale of selected range
<br />Repeatability: <±1.0% of reading3 <±1.0% of reading3
<br />Response Time to Flow: 0.2 sec4 0.2 sec4
<br />Common specifications to all models:
<br />Field Selectable Velocity Ranges
<br />Model 8455/8465: 0.125 m/s to 1.0, 1.25, 1.50, 2.0, 2.5, 3.0, 4.0, 5.0, 7.5, 10.0,
<br />12.5,15.0, 20.0, 25.0, 30.0, 40.0, 50.0 m/s
<br />(25 ft/min to 200, 250, 300, 400, 500, 750, 1000, 1250,
<br />1500, 2000, 2500, 3000, 4000, 5000, 7500, 10000 ft/min)
<br />Model 8475: 0.05 m/s to 0.5, 0.75, 1.00, 1.25, 1.50, 2.0, 2.5 m/s
<br />(10 ft/min to 100, 125,150, 200, 250, 300, 400, 500 ft/min)
<br />Minimum Resolution: 0.07% of selected full scale
<br />Input Power: 11-30 VDC or 18-28 VAC, 350mA max6
<br />Output Impedance:
<br />Voltage mode: less than 1 ohm, 20 mA max source current
<br />Output Resistance:
<br />Current mode: 500 ohms maximum load resistance
<br />Output Signal
<br />(field selectable):
<br />0-5V, 0-10V, 2-10V, 0-20mA, 4-20mA
<br />Output Time Constant
<br />(field selectable):
<br />0.05 to 10 seconds.
<br />Probe Length:
<br />7.5 cm, 15 cm, 22.5 cm or 30 cm (3 in., 6 in., 9 in. or 12 in.)
<br />Temperature Range:
<br />Compensation Range: 0 to 60°C (32 to 140°F)
<br />Electronics and Sensor Operation: 0 to 93°C (32 to 200°F)
<br />Storage : 0 to 93°C (32 to 200°F)
<br />Accessories:
<br />Model 8495 Display, Monitor and Power Supply
<br />*Provides unregulated 12 VDC power (500 mA)
<br />with TSI supplied AC adaptor +[��
<br />*Adjustable full scale reading and signal input 0 �DE
<br />*Digital display indicates velocity
<br />*User selectable audible and visual alarms
<br />Model 8911 Power Supply
<br />*Provides regulated 12 VDC power (500 mA) from 115 VAC source
<br />5.0
<br />N/A
<br />5.0 sec5
<br />12.6 cm
<br />(4.95 in)
<br />Model 8465, 8475
<br />6.4 mm
<br />4.6 mm
<br />(.25 in)
<br />(0.180 in)
<br />7.5 cm (3 in), 15.0 cm (6 in)
<br />3.2 cm
<br />rl"3 io
<br />22.5 cm (9 in), 30.0 cm (12 in)
<br />Model 8455
<br />6.4 mm
<br />(.25 in)
<br />3.4 cm
<br />33 in)
<br />®c
<br />7.5 cm (3 in, 15.0 cm (6 in)
<br />22.5 cm (9 in), 30.0 cm (12 in)
<br />I From 18 to 28°C (64.4 to 82.4°F), outside this range and within temperature compensation range add 0.2% per °C (0.4 per °F).
<br />2 From 20 to 26°C (68 to 78.8°F), outside this range and within temperature compensation range add 0.5%per °C (0.9 per °F).
<br />Directional sensitivity of the Model 8475 is +5%/-20% of reading +0/-0.05 m/s (+0/-10 ft/min) over 270' solid angle regardless of flow direction.
<br />3 Standard deviation, based on one minute average from 0.5 to 5.0 m/s (100 to 1000 ft/min).
<br />4 For 63% of final value, tested at 7.5 m/s (1500 ft/min).
<br />5 For 63% of final value, tested at 2.5 m/s (500 ft/min).
<br />6 Input voltage must be maintained within specifications at the transducer.
<br />7 Calibration occurs with sensor oriented horizontally in a horizontal flow. Uncertainty increases in downward vertical flow at velocities less than 50 ft/min (0.25 m/s).
<br />All specifications subject to change without notice.
<br />TSI Incorporated
<br />Environmental Measurements and Controls Division
<br />500 Cardigan Road
<br />Shoreview, MN 55126 USA
<br />Telephone: 800 777 8356
<br />Europe:
<br />651 490 2711
<br />TSI AB:
<br />Fax: 6514902874
<br />Telephone: 46 18 527000
<br />E-mail: emco@tsi.com
<br />TSI GmbH
<br />Website: www.tsi.com
<br />Telephone: 49 241 523030
<br />PIN 2980139 Copyright O 1999 by TSI Incorporated Printed in U.S.A.