<br /> INTERCONNECT DIAGRAM POWER SPECIF CATIONS � �'��'���
<br /> ������
<br /> ����.
<br /> W►SH A!D WILL.ARCFNIE�15,P.S. NO
<br /> P�IRT FEREAf SFINL�OOPED,
<br /> DUPLICAIED� DIS7R6lRFD, OISCL06ED OR
<br /> P R OT E U S X R/a 8 0 kW oc�ww�r9o�ox���0
<br /> p�Y M1IHORIiED M wRI1RIG BY
<br /> B07E9G1, qASH A!D IIALL AR(�NIEC�IT,
<br /> VOLTAGE PRIMARY SOURCE IS REQUIRED FOR ALL INSTALLATIONS.
<br /> RANGE OF LINE VOLTAGES :
<br /> _ NOMINAL LINE VOLTAGE OF 480, 3 PHASE, WITHOUT NEUTRAL,
<br /> 60 Hz. �
<br /> . REQUIRED POWER SUPPLY: WYE DISTRIBUTION �
<br /> MAXIMUM DAILY VOLTAGE VARIATION MUST FALL WITHIN ONE OF
<br /> THE RANGES IN TABLE A.
<br /> TABLE A CURRENT (AMPS) Z O
<br /> ALLOWABLE NOMINAL NORMAL RANGE MINIMUM STANDARD 0
<br /> I N P U T VOLTAGE t 10 PERCENT MAX. OVERCURRENT �
<br /> VOLTAGES/ MOMENTARY CONTINUOUS PROTECTION
<br /> XTS1 CURRENT 460 414-506 157 5.8 80-A � �
<br /> I pT K D EMAN D
<br /> 480 432-528 151 5.5 80-A
<br /> ALL CALCUI�4TIONS BASED UPON NOMINAL VOLTAGE
<br /> NOTE LOW LINE CONDITIONS MAY INHIBIT SOME HIGH kVp TECHNIQUES. �
<br /> THE GENERATOR AUTOMATICALLY ESTABLISHES THESE INHIBITS
<br /> BASED ON ACTUAL LINE CONDITIONS AND SYSTEM REGULATION.
<br /> PHASE- PHASE-TO-PHASE VOLTAGES MUST BE WITHIN +2 PERCENT � �
<br /> BALANCE OF THE LOWEST PHASE-TO-PHASE VOLTAGE. MAXIMUM ALLOWABLE �
<br /> TRANSIENT VOLTAGE EXCURSIONS ARE 2.5 PERCENT OF RATED = N�
<br /> LINE VOLTAGE AT A MAXIMUM DURATION OF 5 CYCLES AND
<br /> , FREQUENCY OF 10 TIMES PER HOUR. � �N
<br /> 3O' POWER CONTINUOUS POWER DEMAND =4.6 KVA. (MAX DEMAND = 1 25 KVA) � Uj 3�
<br /> DEMAND
<br /> �d � � x
<br /> � o
<br /> Z „ W�
<br /> D U CT TABLE B DEMAND VALUE " V+
<br /> � "C 1 2� "C 1 1 � "C MAXIMUM � � .
<br /> (2)2/ C ) � , ( ) � MOMENTARY kva * 125 � -� � �
<br /> P O W E R POWER FACTOR 0.73 m � ��
<br /> 56� DEMAND. AT � � �o
<br /> ' \,� mA 630 m C[ v��
<br /> X �
<br /> � kVp 80 �N
<br /> 50 aN
<br /> * DEMAND INCLUDES POWER FOR ENTIRE SYSTEM. ��
<br /> LINE VOLTAGE REGULATION AT MAXIMUM POWER DEMAND o 0
<br /> MUST BE LESS THAN OR EQUAL TO 6 PERCENT.
<br /> �a
<br /> 1���C� 3#2,1 �2N, 1 #2G CU ---- - - - _ _ _ DISTRIBUTION F�OR A SINGLE UNIT INSTALLATION, THE MINIMUM TRANSFORMER SIZE � N°�
<br /> D�K 1 TRANSFORMER IS 150 KVA. SYNTHESIZED POWER FEED IS NOT ACCEPTABLE N�"
<br /> � 3/4"C, 2�12, 1 �12G CU
<br /> " 2G CU � - - - - ELECTRICA� NOTES
<br /> 3/4 C, 2�12, 1 �1 XR C 2
<br /> � �,�"C� 2#12, 1 #12 G C U NOTE 1: ALL WIRES SPECIFIED SHALL BE COPPER STRANDED, FLEXIBLE, THERMO-PLASTIC, COLOR CODED, CUT 10 FOOT
<br /> LONG AT OUTLET BOXES, DUCT TERMINATION POINTS OR STUBBED CONDUIT ENDS.
<br /> 2 N 1 B—27 RDS 1 T R ALL CONpUCTORS, POWER, SIGNAL AND GROUND, MUST BE RUN IN A CONDUIT OR DUCT SYSTEM. ELECTRICAL V�
<br /> l CONTRACTOR SHALL RING OUT AND TAG ALL WIRES AT BOTH ENDS. WIRE RUNS MUST BE CONTINUOUS COPPER �
<br /> RX R 1 �N � B—�� STRANDED AND FREE FROM SPLICES. ALUMINUM OR SOLID WIRES ARE NOT ALLOWED. �
<br /> 24VDC 24VDC NOTE 2: WIRE SIZES GIVEN ARE FOR USE OF EQUIPMENT. LARGER SIZES MAY BE REQUIRED BY LOCAL CODES. �
<br /> »
<br /> 1 2��C� 3�2,1 �2 N ,1 #2 G C U �� 3 4 C, 2#12, 1 #12 G C U NOTE 3: ELECTR CA�LMCOD DS D THAT ALL WIRES BE COLOR CODED, AS REQUIRED IN ACCORDANCE WITH NATIONAL AND LOCAL � �
<br /> 5�� g ,8� g ,8' H 0 S P ITAL —� OTE 4: CONDUIT SIZES SHALL BE VERIFIED BY THE ARCHITECT ELECTRICAL ENGINEER OR CONTRACTOR, IN ACCORDANCE WITH v W C
<br /> MA N SW TCHBOARD 4BD �� E�.ECTR CA� � � � 0 V �
<br /> S E E O N E—L �E D AG RAM � LOCAL OR NATIONAL CODES.
<br /> OUT�ET N O T E 5: L O C A T E A T L E A S T O N E C O N V E N I E N C E O U T L E T C L O S E T O T H E S Y S T E M CON T ROL, THE POWER DISTRITBUTION UNIT AND O Z u
<br /> 0 N D R A W I N G E 4. I . (2)2%��C O N E O N E A C H W A L L O F T H E P R O C E D U R E R O O M. U S E H O S P I T A L APPROVED OUTLET OR E QUIVALENT. Q � ,o
<br /> . . Q � �
<br /> � NOTE 6: GENERAL ROOM ILLUMINATION IS NOT ILLUSTRATED. CAUTION SHOULD BE TAKEN TO AVOID EXCESSIVE HEA T FR OM � � �
<br /> OVERHEAD SPOTLIGHTS. DAMAGE CAN OCCUR TO CEILING MOUNTING COMPONENTS AND WIRING IF HIGH WATTAGE BULBS � d
<br /> � ARE USED. RECOMMEND LOW WATTAGE BULBS NO HIGHER THAN 75 WATTS AND USE DIMMER CONTROLS (EXCEPT MR). Y C
<br /> 62 DO NOT MOUNT LIGHTS DIRECTLY ABOVE AREAS WHERE CEILING MOUNTED ACCESSORIES WILL BE PARKED. `W O �
<br /> � PC 9.8' WDR1 �
<br /> � NOTE 7: ROUTING OF CABLE DUCTWORK, CONDUITS> ETC., MUST RUN DIRECT AS POSSIBLE OTHERWISE MAY RESULT IN THE NEED � � �
<br /> , �� ����� � FOR GREATER THAN STANDARD CABLE LENGTHS (REFER TO THE INTERCONNECTION DIAGRAM FOR MAXIMUM USABLE U � �
<br /> LENGTHS POINT TO POINT). �
<br /> 2
<br /> 1���C 1 NOTE 8: CONDUIT TURNS TO HAVE LARGE, SWEEPING BENDS WITH MINIMUM RADIUS IN ACCORDANCE WITH NATIONAL AND LOCAL J
<br /> 1/»C ELECTRICAL CODES. �
<br /> 2 � NOTE 9: A SPECIAL GROUNDING SYSTEM IS REQUIRED IN ALL PROCEDURE ROOMS BY SOME NATIONAL AND LOCAL CODES. IT IS
<br /> 62� RECOMMENDED IN AREAS WHERE PATIENTS MIGHT BE EXAMINED OR TREATED UNDER PRESENT, FUTURE, OR EMERGENCY
<br /> TIB CONDITIONS. CONSULT THE GOVERNING ELECTRICAL CODE AND CONFER WITH APPROPRIATE CUSTOMER ADMINISTRATIVE
<br /> PERSONNEL TO DETERMINE THE AREAS REQUIRING THIS TYPE OF GROUNDING SYSTEM. � �o '�,AS fO�
<br /> NOTE 10: THE MAXIMUM POINT TO POINT DISTANCES ILLUSTRATED ON THIS DRAWING MUST NOT BE EXCEEDED. ti� ? ��
<br /> 1%"C 62' � � x
<br /> NOTE 11: PHYSICAL CONNECTION OF PRIMARY POWER TO GE EQUIPMENT IS TO BE MADE BY CUSTOMERS ELECTRICAL CONTRACTOR M 31 �1 n
<br /> � D WITH THE SUPERVISION OF A GE REPRESENTATIVE. THE GE REPRESENTATIVE WOULD BE REQUIRED TO IDENTIFY THE
<br /> PHYSICAL CONNECTION LOCATION, AND INSURE PROPER HANDLING OF GE EQUIPMENT. c� g��
<br /> NOTE 12: GEHC CONDUCTS POWER AUDITS TO VERIFY QUALITY OF POWER BEING DELIVERED TO THE SYSTEM. THE CUSTOMER'S �sSI�NAL ��'
<br /> D R I M A G I N G 0 P T 0 �l �����, � ELECTRICAL CONTRACTOR IS REQUIRED TO BE AVAILABLE TO SUPPORT THIS ACTIVITY.
<br /> �oe r�e�
<br /> . . . . . . -� n-�
<br /> �►�
<br /> . e-is�n
<br /> � �
<br /> Z �
<br /> _ � �
<br /> � s�r
<br /> �
<br /> . �
<br /> m
<br /> � ��
<br /> �
<br /> N
<br /> �
<br /> ,� QE X-RAY
<br /> N ELECTRICAL
<br /> � DETAILS
<br /> � � FLAG NOTES:
<br /> _
<br /> Q DIAGRAM KE
<br /> W 1�PROVIDE 120V/24V 50VA TRANSFORMER FOR EMERGENCY POWER OFF TO
<br /> � SHUNT TRIP BREAKER FEEDING X-RAY GENERATOR "XGn. - - PROVIDE CONDUIT AND WIRING AS INDICATED.
<br /> m s�r �a
<br /> ' � 2�PROVIDE NORMALLY OPEN RELAY WITH A 24 VOLT, DC COIL PROVIDE RACEWAY FOR GE SUPPLIED CABLWG.
<br /> i AND A STEERING DIODE WIRED OVER THE RELAY COIL TO BE
<br /> �
<br /> ,�, INSTALLED IN J-BOX. SIEMENS TO PROVIDE CONTROL SIGNAL E � ■ �
<br /> 0
<br /> � TO RELAY.
<br /> �-
<br /> 0